Meriam M201 Rotary Gas Meter Tester

M201

Meriam’s M201 has new features to make drop testing Rotary (Natural) Gas Meters even easier! New hardware options protect the M201 from one sided overpressure. Options include a 1/4 turn valve manifold and an integral push-to-read cartridge valve in addition to the traditional push-to-read valve manifold. More importantly, new operating features provide users with a live pressure display and more automation to make field use easier than ever before.

The new Measure Mode lets the M201 be used as a conventional differential pressure measurement device. Users will appreciate the live pressure display as they check the handheld’s zero or observe DP across rotary gas meters. And when a drop test is made, the M201 uses an internal stop watch to measure the test length. The user only has to start and stop the test; the M201 reports both the test time and the average differential pressure measured during the test. Toggle the Min/Max key to see the minimum and maximum DP measured during the test.

This product is an updated version of the 350R.

Features

Accuracy
DN0028: + 0.1% FS 0-28”
DN0200: + 0.05% FS 0-200”
No temperature effect from 23o to 122o F (-5o to +50oC)
Engineering Units
Inches of H2O (select 4oC, 20oC, or 60oF), PSI, mm of Hg, inches of Hg, mBars, Bars, kPa, kg/cm2, cm of H2O and user selectable scaling.
Media Compatibility
DN: Non-isolated differential sensor for clean, dry, non-corrosive gases
Field Re-cal
The M201 can be recalibrated in the field for zero, span, and linearity. The user has the option to select a 1 point or 5 point calibration procedure. Enables users with accurate primary standards to calibrate their devices locally if desired.
Pressure Limits
Twice range on Differential (DN) units when pressurized on high side only and 150 PSI (10.5 Kg/cm2) static when applied to both sides of the sensor simultaneously.
Min/Max Capture
Capture speed is 10X per second
Operation
A push-to-read valve keeps HI and LO sides equalized during connection across meter. Turn M201 “ON” to get message “Press Hold to Acquire Data.” Press “Hold” key to start test. User determines test time, ending test by
pressing the “Hold” key again. Display will show the averaged differential pressure across the meter from test start to test stop. Press the “Hold” key again to start a new test.
Auto Shut-Off
User selectable shut-off. Programmable at 10, 20, 30, 45, 60 minutes and disable
Pressure Ranges
DN0028 0-28” H2O Differential (0.1% F.S.)
DN0200 0-200” H2O Differential (0.05% F.S.)

For More Information

Contact us on 08 8347 0911 or via email below